The Hewlett Packard 1090 HPLC systems are the benchmark for precision, reliability and ease of use for HPLC applications. I can help you with any replacement parts your system may need since I have been repairing and troubleshooting these systems for the last 15 years. With a little time and effort, we can get your system back to a fully-functioning state. Most parts available are for the common autosampler/autoinjector/FPD systems. I don t have a lot for the DAD or manual injector systems (sorry).
This listing is for part QUOTE ONLY. No parts will be sold directly through this listing.
Some common HP1090 consumable / PM parts:
Bottle, system, glass, 1000mL (new or used once)
Metering Pump Seals, for use with buffered solvents (High pH 2.3-12.5), 2 each
Metering Pump sapphire piston, 1 each
Proportioning Valve Capillary Outlet 0.6 mm i.d., 195 mm lg
Check Valve, High Pressure Pump, standard (pH 2.3-9.5), 1 each
Check Valve, High Pressure Pump, (high pH 2.3-12.5), 1 each
Sieve, High Pressure Pump, standard (pH 2.3-9.5), 1 each
Sieve, High Pressure Pump, (high pH 2.3-12.5), 1 each
Oil, High Pressure Pump, 100mL, 1 bottle
Needle Seat, autoinjector, 1 each
Rheodyne Rotor Seal, autoinjector, high PH
Syringe o-ring, Teflon seal 25 l
Syringe Capillary 0.25 mm i.d., 110 mm lg, 1 each
Magazine Assembly, autosampler, 1 each
Deuterium Lamp (Sonntek LH-10)
Some common HP1090 replacement parts and the typical system errors:
Restriction Capillary 0.17 i.d., 500 mm lg (coil of tubing that allows syringe on autoinjector to empty out after sample injected).
Common error: Sample volume variability
Flush Valve (for autoinjector, to allow syringe backflushing).
Common error: Syringe won t backflush
Common error code: EI20, EF20, ED20
Injection Valve Sensor Board (senses when injection valve switches).
Common error code: EI02 and EI10
Cover Switch (for autosampler lid).
STAT Sample Swivel Arm Sensor Board
Common error code: EI04 and EI08
Needle Arm Sensor Board (senses when needle is up or down).
Common error code: EI03 and EI09
Syringe Home Position Sensor Board (senses when syringe is at home position).
Common error code: EI01 and EI011
Bottle in Position Sensor Board (behind autosampler needle, senses when needle is down into sample vial).
Common error code: EI05 and EI07
Autoinjector Sample Loop, Capillary 25 l, 0.6 mm i.d., 1100 mm lg
Common error: Leak at needle connection or at injection valve, both EI20
Autosampler Vial Count Sensor Board (behind needle seat, counts vial positions when needle is up in the air)
Common error code: EI23 (or rack all the way out and grinding )
Oven Door Hinge, 2 each, refurbished, shiny and protected from corrosion, (without screws)
Air Filter (for back of FPD, aftermarket)
I also offer common system components at a reasonable cost. The idea is that if we ve identified the faulty part, you simply remove it from your system, purchase the rebuildable exchange* part, send the faulty part to me to rebuild. I repair, function-test and return the part to you. If your part is broken, corroded/rusted or missing parts, the without exchange pricing would apply. You can always send a photo of your part and I can let you know if it fits the rebuildable exchange* criteria.
Some common HP1090 rebuilt/replacement parts:
Metering pump with rotary valve, 1 each.
$300 each (with rebuildable exchange*)
Metering pump without rotary valve, 1 each.
$200 each (with rebuildable exchange*)
Rotary valve (only) for metering pump, 1 each.
MCD (Metering Channel Driver) Board ( Red Tab Board, controls Metering Pump), 1 each
$150 each (with rebuildable exchange*)
Proportioning valve (MCGV, multi-channel gradient valve assembly), for PV5 system, 1 each
CVD (valve driver) Board for PV5 system ( Purple Tab Board, controls MCGV), 1 each
Low Pressure Compliance (LPC), 1 each
High Pressure Pump Assembly 230 V 50 Hz
Check valves and sieve not included.
$500 each (with rebuildable exchange*)
High Pressure Pump Assembly 115 V 60 Hz
Rheodyne pneumatic injection valve unit
$300 (with rebuildable exchange*),