Beckman coulter flow cytometry epics xl laser system


BECKMAN COULTER FLOW CYTOMETRY EPICS XL LASER SYSTEM
Condition: Pre -owned. Removed from a working enviroment. Powers on and in good cosmetic condition.
Specifications: EPICS XL Flow Cytometry Systems Model Number: XL,mmSerial Number: AE04026Assy No: 6706689Date Of Manufacture: 01/25/2001
A 4-color flow cytometric analyzer with state-of-the-art Digital Signal Processing for reliable linearity and drift free signal amplification and color compensation. System II software helps eliminate tedious manual processes by automating instrument set-up and compensation for 2-, 3- and 4-color applications. It also accommodates a number of optional application systems, including reticulocyte analysis, CD34+ enumeration, DNA analysis and an automated analysis algorithm that allows a full CDC recommended panel in only two tubes. With System II software, EPICS XL systems also provide: AutoPrime Feature If data does not appear within a special set display region, AutoPrime will flush the flow cell as many as three times. If a problem still exists, the system will stop acquisition, ensuring that your samples are not wasted.
Barcode Scanner Provides positive identification of carousel number, tube position and sample, which ensures greater reliability of test results.
FastSet and FastComp Functions Enables users to simply grab a displayed population and place it where they want it the software does the rest.
Detector Parameters/Data Acquisition Up to 8 signals plus Auxiliary, Prism, Time and Ratio can be acquired simultaneously from a possible 21 available parameters Forward Scatter linear, log and peak
Side Scatter linear, log and peak
PMT1-PMT4- linear, log and peak
One Auxiliary (AUX) signal assignable to any linear or peak signal Time
Four Decade logarithmic DIGITAL SIGNAL PROCESSING (DSP) of linear FS, SS, FL1-FL4 data Laser: Uniphase Argon ion, 488nm, 15mW output Flow Cell 250m square channel BioSense enhanced optics quartz mounted with vertical (upwards) flow path for superior hydrodynamic focusing and minimal bubble retention Detectors: Forward Scatter Detector Solid state detector with Patented Fourier design allowing for adjustable sensitivity via Neutral Density Filter
Side Scatter Detector High performance photodiode
Fluorescent Detectors High performance photomultipliers with spectral sensitivity from 200nm-800nm
User Interchangeable Optical Filters.Band Pass (BP): 525nm, 575nm, 620nm, 675nm
Dichroic Long Pass (DL): 488nm, 550nm, 600nm
Resolves 0.5m diameter particles from background noise with maximum detection up to 40m diameter particles Environmental operating temperature: 16-32 degrees C (60-90 degrees F) Dimensions : Allow sufficient space for ventilation, and access for maintenance
30.5 cm (12 in.) from the back
30.5 cm (12 in.) on both sides
Utility Connections Three dedicated 50/60Hz lines: 115VAC, 20A; 220 VAC, 15A